Rapid ultra-performance liquid chromatography assay of losartan potassium in bulk and formulations

Abstract

Background

Losartan potassium is a non-peptide AT1 receptor drug used in the treatment of hypertension.

Methods

A simple, rapid, sensitive, and validated isocratic reverse-phase ultra-performance liquid chromatographic (RP-UPLC) method was developed and validated for the determination of losartan potassium (LOS) in bulk drug and tablets. The assay was developed using Waters Acquity BEH C18 (100 mm × 2.1 mm), 1.7-μm column with a mobile phase consisting of a mixture of phosphate buffer (pH 3.2) and acetonitrile (50:50 v/v).

Results

An assay with a total run time of only 5 min was developed. The method monitored at 245 nm exhibited linearity over a concentration range of 2.0 to 15.0 μg mL−1 LOS. The limits of detection and quantification (signal-to-noise ratio (S/N) = 10) were found be 0.018 and 0.054 μg mL−1, respectively. The intraday and interday RSDs were less than 1.0%. The method was validated by the determination of LOS levels in tablets where the percentage on the label claim was 100 ± 2. The accuracy of the method was further ascertained by recovery studies via the standard addition procedure, which yielded satisfactory results.

Conclusion

A rapid UPLC assay of LOS in bulk drug and tablets was developed and validated.

Background

Losartan potassium, a monopotassium salt of 2-butyl-4-chloro-1-[p-(o-1H-tetrazol-5-ylphenyl)benzyl] imidazole-5-methanol or (2-butyl-4-chloro-1-{[2′-(1H-tetrazol-5-yl)biphenyl-4-yl]methyl}-1H-imidazol-5-yl)methanol (Figure 1), is a non-peptide AT1 receptor antagonist used for the treatment of hypertension (Conlin [2001]). Losartan potassium was the first angiotensin II receptor antagonist to be marketed. Currently, it is marketed by Merck and Co. Inc. under the trade name, Cozaar. Several analytical techniques have been reported for the determination of losartan potassium (LOS) in different pharmaceutical formulations and in biological fluids. Most of these methods employ high-performance liquid chromatography (HPLC) with UV detection (Seburg et al. [2006]; Pedroso et al. [2009]). The drug LOS is determined by HPLC in the presence of other drugs such as hydrochlorothiazide (Carlucci et al. [2000]; Erk [2001]). The LC-MS-MS method works for LOS in the presence of its degradation products and in plasma (Yeung et al. [2000]; Gonzalez et al. [2002]).

Figure 1
figure 1

Molecular structure of the drug, LOS.

The separation science plays a key role in pharmaceutical industry ranging from impurity profiling to the final assay for monitoring the finished products. The batch assay of the pharmaceuticals in a quality control setup needs to be rapid, accurate, and sensitive. The fast quality monitoring of the products using ultra-performance liquid chromatography (UPLC) is gaining pronounced interest. Known for its advanced technology, UPLC is based on the principles of liquid chromatography, which utilizes 1.7-μm column particles. This enhances the separation process without affecting the resolutions. Due to small particle size, the system entails the use of high pressure of the order 15,000 psi to pump the mobile phase. The elevated mobile phase linear velocity results in high resolution, sensitivity, and shorter analysis time. Owing to its speed and sensitivity, this technique is gaining considerable attention in recent years in different fields of pharmaceutical and biomedical analysis (Krishnaiah et al. [2010]; Waren and Tchlitcheff [2006]). The literature survey shows that, despite these advantages, the UPLC has not been applied for the assay of the drug, LOS.

The aim of this study is to develop and validate, according to the current ICH guidelines, a fast, accurate, precise, and sensitive UPLC method for the analysis of LOS in tablets without the interference from inactive ingredients.

Methods

Materials and reagents

The pure active ingredient sample of LOS was given as a gift by Jubilant Life Sciences, Mysore, India. Tablet formulations, viz. Cosart from Cipla, and Covance from Ranbaxy, were purchased from local commercial sources. Solvents such as methanol and acetonitrile (HPLC grade), potassium dihydrogen orthophosphate, and orthophosphoric acid (Qualigens, Mumbai, India) were purchased from Merck & Co., Inc., Whitehouse Station, NJ, USA. Double-distilled water was used throughout the investigation.

Chromatographic conditions and equipment

Analyses were carried out on a Waters Acquity UPLC (LabX, Midland, ON, Canada) with a tunable UV detector. The output signal was monitored and processed using an Empower software. The chromatographic column used was Waters Acquity UPLC BEH C18 (100 mm × 2.1 mm with 1.7-μm particle size). The isocratic elution process was adopted throughout the analysis.

Mobile phase preparation

An aqueous phosphate buffer solution was prepared by dissolving 2.000 g of potassium dihydrogen orthophosphate in approximately 400 mL of water. The pH of this solution was adjusted to 3.2 with 10% phosphoric acid before diluting to a final volume of 500.0 mL. A mixture of 500.0 mL phosphate buffer and 500.0 mL acetonitrile was stirred and filtered using a 0.22-μm nylon membrane filter. This solution was also used as the solvent in all the subsequent preparations of analyte samples.

Instrumental parameters

The isocratic flow rate of the mobile phase was maintained at 0.20 mL min−1. The column temperature was adjusted to 40°C. The injection volume was 4.0 μL. The elution was monitored at 220 nm and the total run time was 5.0 min.

Preparation of stock solution

A standard stock solution of LOS (100 μg mL−1) was prepared by dissolving an accurately weighed amount (5.0 mg) of the drug in the solvent to a final volume of 50.0 mL in a volumetric flask.

Procedures

Procedure for calibration curve

Working standard solutions containing 2.0 to 15.0 μg mL−1 of LOS were prepared by serial dilutions of the stock solution (100 μg mL−1). Aliquots of 4.0 μL were injected (six injections) and eluted with the mobile phase under the reported chromatographic conditions. The average peak area vs. concentration of LOS (in μg mL−1) was plotted to obtain the standard curve for determining the unknown concentrations of the analyte. Alternatively, the corresponding regression equation was derived using mean peak area-concentration data, and the concentration of the unknown analyte can be computed from the regression equation.

Preparation of tablet extracts and assay procedure

Ten tablets (approximately 25-mg LOS each) were weighed and ground to a fine powder. A quantity equivalent to a tablet was weighed and transferred into a 100-mL flask, dissolved in 60 mL of the mobile phase, sonicated for 20 min, and then diluted to the final volume with the same mobile phase to yield a concentration of 250 μg mL−1. The solution was passed through a 0.22-μm nylon membrane filter. Appropriate volumes of the filtered solution were diluted with the mobile phase to obtain the desired concentrations such as 5.0 ppm.

Procedure for method validation

Accuracy and precision

To determine the accuracy and intraday precision, pure LOS solutions of three different concentrations were analyzed in six replicates each during the same day. Mobile phase was injected as the blank solution before the sample injection, and the relative standard deviation (%RSD) values of the peak area and retention time were determined.

Limits of detection and quantification

The limit of detection (LOD) and limit of quantification (LOQ) were determined by the signal-to-noise (S/N) ratio method. These were obtained by a series of dilutions of the LOS stock solution. Precision study was performed at the LOQ level also. LOQ solution was injected six times (n = 6) and the %RSD values for the obtained peak area and retention time were calculated.

Solution and mobile phase stability

The stability of the LOS solution was tested by injecting the sample into the C-18 column. The peak area was recorded at time intervals of 0, 3, 6, 8, 12, and 15 h, and the RSD and time values were calculated. The mobile phase stability was studied by injecting a freshly prepared sample solution at the same time periods, and the RSD values of the peak areas were calculated. The RSD values of both studies were found to be less than 3.0% exhibiting compatibility of the diluent and the stability of the mobile phase.

Results and discussion

Method development

The drug LOS is a basic due to the presence of imidazole and tetrazole moieties in the molecule (Figure 1). The log P value of LOS was found to be 2.25 by ChemDraw Ultra Version 7.0 indicating a high lipophilicity. Acidic mobile phase with a buffer of pH 3.2 was chosen to protonate nitrogen atoms for a fast elution. In order to achieve better efficiency of the chromatographic system, the experimental conditions such as composition and pH of the mobile phase, detection wavelength, nature of column, and column temperature were optimized by varying one parameter at a time while keeping the other conditions constant. Several proportions of buffer, water-acetonitrile mixture, and methanol were evaluated to obtain suitable composition of the mobile phase. Parameters such as the retention time, peak shape, theoretical plates, and run time were the major tasks while developing the method. Several combinations of gradient methods were also performed. Isocratic method was found to be better for the assay. Finally, at the mobile phase composition described under the ‘Methods’ section, the method gave the lowest peak tailing factor and the highest theoretical plate count.

Validation of the method

The described method for the assay of LOS has been validated as per the current ICH Q2 (R1) guidelines.

Analytical parameters

Figure 2 shows an overlay of the chromatographic peaks for LOS at different concentrations. A linear correlation was obtained between the peak area and the concentrations in the range 2.00 to 15.0 μg mL−1 LOS, from which the linear regression equation was computed to be A = 96,192.3C − 3,271.6 (r = 0.9999), where A is the peak area and C is the concentration of LOS (in μg mL−1). The LOD and LOQ values, slope, y-intercept, and their standard deviations evaluated are presented in Table 1 and Figure 3. These results confirm the existence of a linear relationship between the concentrations of LOS and the peak areas along with the sensitivity of the method.

Figure 2
figure 2

Overlay of chromatograms for losartan with concentrations (a) 2 μg mL −1 , (b) 5 μg mL −1 , (c) 8 μg mL −1 , (d) 10 μg mL −1 , (e) 12 μg mL −1 , and (f) 15 μg mL −1 .

Table 1 Regression and sensitivity parameters for UPLC analysis of LOS
Figure 3
figure 3

Linear calibration plot of peak area vs. LOS concentration.

Accuracy and precision

Table 2 shows the data for three concentration levels of the analyte LOS in tablets determined with and without the standard addition of an equal amount of the pure LOS. The data include method accuracy and precision represented by the relative error (RE) and RSD, respectively, for six determinations at each concentration level. Both RE and RSD decrease with the increase in the sample size showing the lowest values for the 8 μg mL−1 sample size. However, the relatively low %RSDs indicate the high precision of the method.

Table 2 Results of determination of LOS in tablets and in standard samples

Robustness of the method and stability of the solution

The robustness of an analytical procedure is a measure of its capacity to remain unaffected by small but deliberate variations made in method parameters. It provides an indication of its reliability during normal usage. At varied chromatographic conditions (flow rate, temperature, and mobile phase composition, absorption wavelength of the detector, etc.), the analyte peak %RSD, tailing factor, and theoretical plate count should remain close to the actual values. The RSD values ranging from 0.06% to 0.13% represent the robustness of the proposed system. Results of RE and RSD determined for the variation of temperature and absorption wavelength are presented in Table 3. At a specified time interval, values of %RSD for the peak areas obtained for the stabilities of the drug solution and the mobile phase were within 1%. This shows that there is no significant change in the elution of the peak and its system suitability criteria (%RSD, tailing factor, theoretical plate count). The results also confirm that the standard drug solution in the mobile phase was stable at least for 15 h during the assay performance. The constant peak area for different time intervals is shown in Figure 4.

Table 3 Robustness of the analytical method
Figure 4
figure 4

Chromatograms for LOS solution stability for periods of 3, 6, 8, 12, and 15 h in triplicate.

Selectivity

Selectivity of the method was evaluated by injecting the mobile phase, standard drug solution, and tablet extract. Under the experimental conditions, there was only a single peak for the drug (LOS) elusion, which showed a high selectivity of the method.

Application to tablet

Aliquots containing 250 μg mL−1 of LOS in extract and the pure drug solution were separately injected in triplicate to the UPLC system. The mean peak area of the tablets was found to be equivalent to that of the pure drug used as a positive control at the same concentration level.

Recovery study

A standard addition procedure was used to evaluate the accuracy and precision of the method. The solutions were prepared by spiking the pure LOS solution into a pre-analyzed LOS tablet extract at three different concentration levels followed by injection into the chromatographic column. The relatively low values of %RE and %RSD show high accuracy and precision of the developed UPLC method for the drug, LOS.

Conclusions

A rapid, isocratic reverse-phase ultra-performance liquid chromatographic (RP-UPLC) method was developed for the quantitative analysis of the drug, losartan potassium or LOS, in pharmaceutical dosage formulations. This method is precise, accurate, robust, and specific. Satisfactory results were obtained from the validation of the method. The short retention time obtained (2.40 min) enables rapid determination of LOS, which is important for its routine analysis in quality control. The method exhibits an excellent performance in terms of sensitivity and speed. Each experiment was repeated three times. Recovery of known amounts of the added analyte was calculated and is presented in Table 2. The reported values of the recovery are the average of three experiments.

References

  1. Carlucci G, Palumbo G, Mazzeo P, Quaglia MG: Simultaneous determination of losartan and hydrochlorothiazide in tablets by high-performance liquid chromatography. J. Pharm Biomed Anal 2000, 23: 185–189. 10.1016/S0731-7085(00)00268-5

    Article  CAS  Google Scholar 

  2. Conlin PR: Efficacy and safety of angiotensin receptor blockers: a review of losartan in essential hypertension. Curr Ther Res Clin Exp 2001, 62: 673–673. 10.1016/S0011-393X(01)80019-9

    Article  CAS  Google Scholar 

  3. Erk N: Analysis of binary mixtures of losartan potassium and hydrochlorothiazide by using high performance liquid chromatography, ratio derivative spectrophotometric and compensation technique. J Pharm Biomed Anal 2001, 24: 603–611. 10.1016/S0731-7085(00)00434-9

    Article  CAS  Google Scholar 

  4. Gonzalez L, Lopez JA, Alonso RM, Jimenez RM: Fast screening method for the determination of angiotensin II receptor antagonists in human plasma by high-performance liquid chromatography with fluorimetric detection. J Chromatogr A 2002, 949: 49–60. 10.1016/S0021-9673(01)01496-0

    Article  CAS  Google Scholar 

  5. Krishnaiah C, Reddy AR, Kumar R, Mukkanti K: Stability-indicating UPLC method for determination of Valsartan and their degradation products in active pharmaceutical ingredient and pharmaceutical dosage forms. J Pharm Biomed Anal 2010, 53: 483–489. 10.1016/j.jpba.2010.05.022

    Article  CAS  Google Scholar 

  6. Pedroso CF, de Oliveira JG, Campos FR, Gonzalves AG, Trindade CLB, Pontarolo R: A Validated RP–LC Method for Simultaneous Determination of Losartan Potassium and Amlodipine Besilate in Pharmaceutical Preparations. Chromatographia 2009, 69: S201-S206. 10.1365/s10337-009-1005-3

    Article  CAS  Google Scholar 

  7. Seburg RA, Ballard JM, Hwang TL, Sullivan CM: Photosensitized degradation of losartan potassium in an extemporaneous suspension formulation. J Pharm Biomed Anal 2006, 42: 411–422. 10.1016/j.jpba.2006.04.030

    Article  CAS  Google Scholar 

  8. Waren SAC, Tchlitcheff P: Use of ultra-performance liquid chromatography in pharmaceutical development. J Chromatogr A 2006, 1119: 140–146. 10.1016/j.chroma.2006.02.052

    Article  Google Scholar 

  9. Yeung PK, Jamieson A, Smith GJ, Fice D, Pollak PT: Determination of plasma concentrations of losartan in patients by HPLC using solid phase extraction and UV detection. Int J Pharm 2000, 204: 17–22. 10.1016/S0378-5173(00)00453-1

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding authors

Correspondence to Tarab J Ahmad or Bellale M Venkatesha.

Additional information

Competing interests

The author declares no competing interests.

Authors’ original submitted files for images

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ahmad, T.J., Raj, A., Radhika, R.T. et al. Rapid ultra-performance liquid chromatography assay of losartan potassium in bulk and formulations. J Anal Sci Technol 5, 33 (2014). https://doi.org/10.1186/s40543-014-0033-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40543-014-0033-2

Keywords