Analysis of intensity and sensitivity of single- and multiple-channel RF head coilsin 3.0-T MRI system

Abstract

Background

High-resolution anatomical and functional images can be acquired using high- andultrahigh-field magnetic resonance imaging (MRI) systems. An increase in the mainmagnetic field strength results in high resolution but suffers from thedisadvantage of field nonuniformity in the radio frequency (RF). To overcome thisRF field inhomogeneity, parallel RF transmission system and sensitivity encodingare widely used.

Findings

Our experimental results showed that the best signal intensity was in transmit andreceive mode head coil, except for the T2-weighted fast field echo (FFE). The bestsignal-to-noise ratio (SNR) was in the 32-channel head coil.

Conclusion

In general, multiple-channel coils have been known as more efficient materials interms of SNR. However, they may not be as important in experiments where intensityis an important factor. Therefore, a suitable coil should be selected using apulse sequence.

Findings

Introduction

In clinical and research fields, 3.0-T magnetic resonance imaging (MRI) systems havebeen widely used. Increasing the main magnetic field strength (≧3.0 T)offers the advantage of high resolution but results in the disadvantage ofnonuniformity in the radio frequency (RF) field because of the short RF wave length(Hoult and Phil [2000]). To exploit the inherent advantages and to overcome the challenges ofusing high-field scanners for the MRI, dedicated RF coils such as multi-channel RF(Adriany et al. [2010]), sensitivity encoding (Pruessmann et al. [1999]), transverse electromagnetic (Vaughan et al. [1994]), and micro-strip transmission line volume coils (Zhang et al. [2001]) have been developed.

In the present study, four RF head coils [transmit and receive mode (T/R) head coiland 8-, 16-, and 32-channel head coils] were evaluated and compared in a 3.0-T MRIsystem. The commercial RF head coils were provided by Philips Medical System(Cleveland, OH, USA). Except for the T/R head coil, the multi-channel RF coilsoperated in the Rx mode only. For high-quality image acquisition, an MRI user had toadaptively select the RF coil according to a given RF pulse sequence. Each RF coilperformance was evaluated and compared in terms of signal-to-noise ratio (SNR), sliceprofile, and RF field uniformity.

As a result, a parallel RF transmission system and an increased number of RF coilchannels can improve the SNR. However, the T/R head coil has the highest signalintensity with a high noise level. An MRI user must adaptively select the RF coil forthe image using the RF coil evaluation results in this study.

Availability and requirements

Methods

The purpose of this study is to evaluate the T/R head coil and the 8-, 16-, and32-channel head coils using each RF pulse sequence [T1-weighted spin echo (SE),T1-weighted fast field echo (FFE), T2-weighted turbo SE (TSE), and T2-weighted FFE]in terms of SNR, slice profile of the images, and RF field uniformity. The imagepositions were at the center and edge (slice offset, 40 mm) slices of the waterphantom. These evaluation and comparison studies were conducted on a Philips Achieva3.0-T MRI system (at the Korea Basic Science Institute, Ochang, Republic of Korea)using a uniform phantom provided by Philips Medical System.

MR pulse sequence

Two types of magnetic resonance (MR) pulse sequences were tested: the SE RF pulsesequence and the FE RF pulse sequence which is the same as the gradient echo.

Figure 1 shows the basic MR pulse sequences of the SEand FE.

Figure 1
figure 1

SE and FE pulse sequence diagrams. Left, SE; right, FE.

Experimental conditions

The images were acquired by the Philips Achieva 3.0-T MRI system. The T/R headcoil was operated in the Tx/Rx mode in the quadrature transmission mode, and theother multi-channel coils used the Rx mode only. Two images (center and edgeslices with a 40-mm gap in the water phantom) were acquired per type of pulsesequence.

The parameters for the T1-weighted images with the SE pulse sequence are echo time(TE)/repetition time (TR) = 10/500 ms, field of view(FOV) = 240 × 240 mm2, reconmatrix size = 560 × 560, slicethickness = 5 mm, and flip angle = 90°.

The parameters for the T2-weighted images with the SE pulse sequence areTE/TR = 80/3,000 ms,FOV = 240 × 240 mm2, recon matrixsize = 560 × 560, slicethickness = 5 mm, flip angle = 90°, refocusingangle = 120° (refocusing control), and TSEfactor = 15.

The parameters for the T1-weighted images with the FFE pulse sequence areTE/TR = 4.60414696 (in-phase)/250 ms,FOV = 240 × 240 mm2, recon matrixsize = 560 × 560, slicethickness = 5 mm, and flip angle = 80°.

The parameters for the T2-weighted images with the FFE pulse sequence areTE/TR = 16.1145134 (in-phase)/685 ms,FOV = 240 × 240 mm2, recon matrixsize = 560 × 560, slicethickness = 5 mm, and flip angle = 18°.

The MR images were acquired using a water phantom (16-cm-diameter refill bottleprovided by Philips Medical System). Four RF coils were evaluated using the fourtypes of sequences (SE and FE with T1 and T2 weighting).

All experiments were performed under the same condition. The signal-scale rangeswere set to be the same prior to the image acquisition. After the imageacquisition, the contrast and signal-scale ranges were set to be the same usingMATLAB (Mathworks, Natick, MA, USA).

Results

The digital imaging and communications in medicine (DICOM) images of the edge andcenter slices are shown in Figures 2 and 3, respectively. They were set at the same signal scale and contrastranges. From the results, the T/R head coil shows maximum image signal intensitylevel in all sequence parameters except for the T2-weighted FFE.

Figure 2
figure 2

MR images obtained by different types of pulse sequences and RF coils.The DICOM images have the same intensity scale (edge slice in waterphantom-foot direction).

Figure 3
figure 3

MR images obtained by different types of pulse sequences and RF coils.The DICOM images have the same intensity scale (center slice in waterphantom).

Increasing the RF coil elements resulted in the decrease in the signal intensity. Onthe other hand, the 32-channel head coil has the highest SNR and the lowest noiselevel, as shown in Figures 4 and 5.

Figure 4
figure 4

Slice signal intensity profile of the center slice images.

Figure 5
figure 5

Slice signal intensity profile of the edge slice images.

Additionally, we confirmed that the T/R head coil has the highest signal intensity,and the multi-channel RF coils have a high SNR. Figure 6shows the positions of each image where the signal intensity, noise (three-pointcenter slice), and SNR were calculated.

Figure 6
figure 6

Calculated positions of the signal intensity and noise.

Conclusions

With the exception of the T2-weighted FFE, the T/R head coil has the advantage interms of signal intensity. On the other hand, the 32-channel head coil shows the bestperformance in terms of SNR.

In the T2-weighted FFE imaging, we can use an 8-channel head coil for the signalintensity and a 32-channel head coil for SNR.

For the other weighted images, a T/R head coil can be selected for the signalintensity, and a 32-channel head coil can be selected for SNR.

In general, multiple-channel coils have been known to be more efficient materials interms of SNR.

However, they may not be as important experiments where the signal intensity is animportant factor. Therefore, a suitable coil should be selected using the pulsesequence.

In other words, the coil that must be used in experiments that require high intensity(such as fMRI) and a signal from a phantom or a coil in terms of its purpose (such asSNR and CNR) in the anatomical image should be selected carefully.

References

  1. Adriany G, Auerbach EJ, Snyder CJ, Gözübüyük A, Moeller S, Ritter J, Van de Moortele PF, Vaughan T, Uğurbil K: A 32-channel lattice transmission line array for parallel transmit and receive MRIat 7 tesla. Magn Reson Med 2010, 63(6):1478–1485. 10.1002/mrm.22413

    Article  Google Scholar 

  2. Hoult DI, Phil D: Sensitivity and power deposition in a high-field imaging experiment. J Magn Reson Imaging 2000, 12(1):46–67. 10.1002/1522-2586(200007)12:1<46::AID-JMRI6>3.0.CO;2-D

    Article  CAS  Google Scholar 

  3. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P (1999) SENSE: sensitivity encoding for fast MRI. Magn Reson Med 42: 952–962

    Article  CAS  Google Scholar 

  4. Vaughan JT, Hetherington HP, Otu JO, Pan JW, Pohost GM: High frequency volume coils for clinical NMR imaging and spectroscopy. Magn Reson Med 1994, 32(2):206–218. 10.1002/mrm.1910320209

    Article  CAS  Google Scholar 

  5. Zhang X, Ugurbil K, Chen W: Microstrip RF surface coil design for extremely high-field MRI andspectroscopy. Magn Reson Med 2001, 46(3):443–450. 10.1002/mrm.1212

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Junyong Park.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

JP designed experiments and wrote the manuscript. JC analyzed the data and CL supervisedthe study. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (https://creativecommons.org/licenses/by/4.0), which permits use, duplication, adaptation, distribution, and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Cite this article

Park, J., Cho, J. & Lee, C. Analysis of intensity and sensitivity of single- and multiple-channel RF head coilsin 3.0-T MRI system. J Anal Sci Technol 5, 25 (2014). https://doi.org/10.1186/s40543-014-0025-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40543-014-0025-2

Keywords