Method development for simultaneous detection of ferulic acid and vanillin using high-performance thin layer chromatography

Abstract

Background

A simple, accurate, and reliable high-performance thin-layer chromatography (HPTLC) method was developed for separation and detection of ferulic acid and vanillin.

Methods

Separation of ferulic acid and vanillin was carried out on 20 × 10 cm thin layer chromatography (TLC) plates using mobile phase containing toluene/1, 4-dioxan/acetic acid in the ratio 9:2.5:0.4 (v/v). The FA and vanillin were scanned at 320 and 312 nm, respectively. Method was validated for linearity, accuracy, precision, robustness, limit of detection, limit of quantification, and specificity.

Results

Retention factor (Rf) obtained for ferulic acid and vanillin was 0.48 and 0.56, respectively. The correlation coefficients, 0.9975 and 0.9991 with an average recovery of 98.77% and 98.45% obtained for ferulic acid and vanillin respectively by this method were satisfactory.

Conclusion

The optimized method was found to be efficient, precise, accurate, specific, and economic. Therefore, the method would be useful for both qualitative and quantitative routine analysis in pharmaceutical, food industry, and research laboratories.

Background

Aromatic compounds are present in natural sources with substantial combinations which are directly responsible for its odor and sensitivity. They can be categorized as volatile organic compounds like aldehydes, alcohols, ketones, esters, lactones, and terpenes (Raisi et al. [2008]). They are known to be precursors for the production of numerous products employed in the food, pharmaceutical, and chemical industries and are present at very low concentrations in natural sources. According to US and European legislations, synthetic flavor production is not considered as natural. Alternatively, biotechnology offers microorganisms as production hosts for different types of aromatic compounds in industrial fermentative processes (Lomascolo et al. [1999]). The most intensively studied biotransformation using microorganisms is the bio conversion of ferulic acid (FA) to produce natural vanillin (Priefert et al. [2001]).

FA is an important precursor of vanillin that is available in abundance in plant cell walls linked to polysaccharide by an ester or ether bonds (Xu et al. [2005]). FA is a potent antioxidant because it effectively scavenges free radicals and even possesses antimicrobial properties by preventing the lipid peroxidation caused by microbes (Graf [1992]). Moreover, it is used in cosmetics for the photo protection of skin and in protection against various inflammatory diseases.

Vanillin is widely used in food industry as a flavoring agent but also has applications in some fragrances and pharmaceuticals (Priefert et al. [2001]). It is also known to possess anti-metastatic, anticancer (Ho et al. [2009]) and anti-inflammatory (Wu et al. [2009]) activities. It exhibits antimicrobial properties due to its phenolic nature and hence used to develop antimicrobial films used in packaging of bakery products (Rakchoy et al. [2009]).

FA and vanillin are generally determined by various chromatographic methods. Different approaches such as UV spectrophotometry (Mabry et al. [1970]; Macheix et al. [1990]) gas chromatography (GC), capillary electrophoresis (CE), high-pressure liquid chromatography (HPLC), thin layer chromatography (TLC), and high-performance thin layer chromatography (HPTLC) are some of the frequently used methods for the detection, qualitative analysis, and quantification (Sharma et al. [2007]).

Spectrophotometric methods are used for identification of phenolic acids and are generally carried between a range of 220 to 320 nm (Mabry et al. [1970]; Macheix et al. [1990]); however, methods such as the Folin Ciocalteu spectrophotometric method results in nonspecific detection of the phenolic compounds and the interference of components such as ascorbic acid in food samples, that behave as reducing agents. Absorption of phenolic compounds is affected by pH, solvents used in the method and the interference of proteins and amino acids (Constantine et al. [2007]).

Volatile compounds are directly analyzed by gas chromatography, a technique of unsurpassed separation capacity (Sostaric et al. [2000]). GC is a major chromatographic technique employed for the analysis of essential phenolic acids in plants. It deals with high sensitivity and selectivity (Chiou et al. [2007]) but requires derivatization step of hydroxyl groups in phenolic compounds. They are modified by various reagents to make more volatile compounds by a process such as methylation, conversion into trimethylsilyl (TMS) derivatives. However, problems such as poor separation and low stability after derivatization state are some of the shortcomings of this method. CE is too employed for analysis of phenolic compounds (Huck et al. [2005]; Butehorn et al. [1996]). Mostly, the method falls in the field of natural product research, including the analysis of plants, vegetables, herbs, and other plant- or fruit-derived products. It results in oxidation of phenolic compounds by dissolved oxygen and increase in migration time of flavonoids due to the increase in buffer concentrations (Constantine et al. [2007]).

TLC methods have the ability to screen phenolic compounds easily (Tilay et al. [2008]). The results obtained by TLC method are generally quantified using more multifaceted techniques like HPTLC (Mabinya et al. [2006]). However detection of vanillin by spraying with 2, 4-dinitrophenylhydrazine (2, 4-DNPH) is not significant as the peaks are not detected properly. Currently, the main qualitative and quantitative techniques for phenolic compound detection are HPLC (Rao et al. [1999]; Zheng et al. [2007]). The European pharmacopoeia suggests the development of such analytical method which demands the adequate amount of reagents, solvents, and material (European Pharmacopoeia [2008]). HPTLC allows for the simultaneous analysis of large sample size using small quantities of solvents, thus reducing time and cost of the analysis. The sensitivity for phenolic compounds performed by HPTLC is more as compared to HPLC (Prinjaporn et al. [2013]). Mobile phase having pH 8 and above can be employed. Sample with turbidity and different combinations of solvent can be directly applied. It facilitates automated application and repeated scanning of the chromatogram with the same or different parameters (Bakshi et al. [2002]). Therefore, this technique should be taken into consideration as an alternative to HPLC.

HPTLC is a sophisticated instrumental technique which allows a fast and inexpensive method for analysis. Special advantage of HPTLC includes high sample throughput and low cost per analysis. HPTLC offers a great variety of stationary phases with unique selectivity for mixture components and their separation simultaneously. Processing of standards and samples identically on the same plate directs to better accuracy and precision of method for assessment. HPTLC development is extensive as the mobile phases are fully evaporated before the detection step thus preventing solvent interference in analysis. It minimizes exposure risks and significantly discarding toxic organic effluent problems were reduced thereby reducing possibilities of environment pollution. In response to this, HPTLC-based methods could be considered as a good alternative as they are being explored as an important tool in routine analysis. The HPTLC-developed method is actively used in application of qualitative and quantitative analyses of a wide range of compounds, such as herbal and botanical dietary supplements and nutraceuticals. It helps in identifying compounds present in a given substance; to check starting raw materials (plant extracts, extracts of animal origin, fermentation mixtures) identification of drugs and their metabolites in biological media such as urine, plasma, or gastric fluid (pharmacological, toxicological, pharmacokinetic) (Renger [1993]; [1998]).

The aim of the present study is to develop rapid, economic, selective HPTLC method for analysis and simultaneous determination of FA and vanillin with proper peak separation and hence can be used for routine high-throughput detection and determination of phenolic compounds. It is a one-step biotransformation process using culture Pycnoporus cinnabarinus which undergoes propenoic acid chain degradation of FA to vanillin (Tilay et al. [2008]).

Methods

Chemicals/Reagents

Standard FA and vanillin (99% purity), maltose, diammonium tartarate, yeast extract, malt extract, potassium dihydrogen orthophosphate, magnesium sulfate, calcium chloride, thiamine hydrochloride, sodium hydroxide, and agar powder were procured from Hi-Media Laboratory (Mumbai, India). HPLC grade toluene, 1, 4-dioxan, acetic acid, ethyl acetate, and methanol were procured from Hi-Media (Mumbai, India).

Instrumentation

CAMAG TLC system consist of a CAMAG Linomat V (Muttenz, Switzerland) sample applicator and CAMAG TLC Scanner 3 controlled by WinCats software (1.4.3.6336); CAMAG glass twin-trough chambers (20 × 10 × 4 cm3); 100-μl Hamilton syringe; Silica gel plates (G60F254, 20 × 10 cm) were procured from E-Merck Pvt. Ltd., Mumbai, India. For extraction of vanillin, Buchi evaporator system was used consisting of Buchi evaporator R-124, Buchi water bath B-480, and Buchi vacuum controller B-721.

Preparation of stock solution

Standard stock solutions were prepared by dissolving 25 mg of vanillin and 25 mg of FA in 25 ml of methanol. Working standard solution was diluted (1:200) from stock solution of FA and vanillin to attain concentration of 5 μg/ml.

Microorganism

Pycnoporus cinnabarinus NCIM 1181 was procured from National Centre for Industrial Microorganism (NCIM), National Chemical Laboratory (NCL), Pune, India. Culture was maintained on potato dextrose agar slants at 4°C.

Media for vanillin production

A production media consisting of maltose 20 g/l, diammonium tartarate 1.8415 g/l, KH2PO4 0.2 g/l, CaCl2 2H2O 0.0132 g/l, MgSO4·7H2O 0.5 g/l, yeast extract 0.5 g/l, and thiamine hydrochloride 2.5 mg/l was adjusted to pH 7 and inoculated using mycelium fragments of P. cinnabarinus (Gross et al. [1993]). This was further incubated on an incubator shaker (150 rpm) at 37°C for 6 days. After 3 days of growth, sterile solution of FA (0.03 g/100 ml) was added to the media prepared by dissolving FA in 0.1 N NaOH in the form of sodium ferulate.

Extraction of FA and vanillin from the culture media

The broth obtained after fermentation was filtered and the filtrate was acidified to obtain a pH of 1 to 2.80 ml of the above acidified solution was extracted thrice with equal volume of ethyl acetate. The extracts and the residue were redissolved in 50% (v/v), respectively. The organic phase containing FA and vanillin was concentrated (up to 2 to 3 ml) using a rotary vacuum evaporator with conditions (55°C, 150 rpm, <80 mbar) followed by reconstitution in 2 ml of methanol 50% (v/v) (Tilay et al. [2010]). This solution was used for quantification of FA and vanillin by HPTLC.

Instrumentation and chromatographic parameters

HPTLC was executed using silica gel 60F254 plates. A 10 μl of standard working solutions of FA and vanillin were applied to the plates of size 20 × 10 cm with a 5-mm band length. Ascending chromatography silica gel plate development traveled to a distance of 85 mm at temperature of 25°C with toluene/1, 4-dioxan/acetic acid (9:2.5:0.4; v/v) as mobile phase. After development, the plates were dried and chromatograms were recorded at 320 and 312 nm, respectively using CAMAG TLC Scanner 3. Quantitative evaluation was performed with WinCats software using deuterium lamp having slit width of 5 × 0.45 mm and an application rate of 150 nL/s.

Validation of method

Validation was performed in terms of linearity, specificity, precision, limit of detection (LOD), limit of quantification (LOQ), robustness, and system suitability by ICH guidelines (CPMP/ICH/381/95).

Linearity

Standard solution of FA (50 to 250 ng/band) and vanillin (50 to 200 ng/band) with varying volumes were applied on TLC plates. A plot of peak area against concentrations and its respective standard deviation (SD) and coefficient of correlation was calculated for both the compounds.

Precision

Precision specifies random errors. Results were expressed in percent relative standard deviation (%RSD < 2). Standard solution of FA (100, 150, and 200 ng/band) and of vanillin (100, 150, and 200 ng/band) were applied. Intra-day precision was evaluated by applying each concentration five times on the same day. Inter-day precision was evaluated by applying each concentration five times on three different days with an interval of 24 h.

Recovery

The recovery was used to determine the accuracy of the method. Recovery of FA and vanillin with three different concentration namely, 50, 100, and 150 ng (n = 3) was performed. Samples after extraction were spiked with standard concentration and applied. Peak areas of standard added to samples were calculated and average percent recovery was estimated. Further average percent recovery was calculated.

Specificity

Specificity was performed to compare the standard FA and vanillin and extract. It was calculated by comparing the Rf of the peak, peak start, peak apex, and peak end of the standard and extract. The spectral scan of both the standard and extract was compared. Spectrum scan accelerated at 100 nm/s with split dimensions 5 × 0.45 mm, micro. Spectral detection for optimum wavelength was calculated in the range 200 to 700 nm.

Robustness

Modifications in mobile phase concentration, mobile phase volume, mobile phase saturation time, and temperature were examined to check the robustness. Standard solutions of FA (150 ng/band) and vanillin (150 ng/band) were applied thrice and %RSD of each compound was calculated.

Limit of detection and limit of quantification

LOD is the lowest amount of compound that can be detected with signal-to-noise ratio of 3:1 and LOQ is the lowest amount of compound which can be quantified by signal-to-noise ratio 10:1 with adequate precision and specificity. LOD and LOQ of FA and vanillin were calculated.

System suitability

System suitability was executed to check the reproducibility and resolution of the method. Standard solution of both the standards of 150 ng concentration (n = 5) was analyzed on the same chromatographic plates. After development, plates were scanned and the peak area of each concentration and their Rf values were calculated.

Results and discussion

A GC method for detection of phenolic compounds is complicated and unsuitable for rapidly analyzing profuse samples and with derivatization step. HPLC-ECD, a new method was developed for the analysis of phenolic compound. However, use of HPLC with ECD is not feasible as a routine analytical method (Takahashi et al. [2013]). Even with the HPLC UV detector system, the analysis is time consuming and the quantity of the solvents is the main concern. In the present study, two compounds FA and its bio-transformed product vanillin were quantified using HPTLC. Various mobile phases were screened for proper separation of both the compounds. Well-defined peaks with Rf values 0.48 for FA and 0.56 for vanillin were obtained using toluene/1, 4-dioxan/acetic acid (9:2.5:0.4; v/v) (Olsson et al. 1974). Experimental conditions should be selected when the compounds to be analyzed moves towards or near to the center of the layer (Rf = 0.5) (Srivastava [2011]). Krishna veni et al. ([2013]) used methanol/water/glacial acetic acid (20:5:2; v/v) as the mobile phase and resulted with Rf 0.84 for vanillin. For better peak detection and Rf, the polarity can be reduced. Other mobile phases were compared which resulted in tailing of peaks (Sharma et al. [2007]) using methanol/water/isopropanol/acetic acid (30:65:2:3; v/v) and inappropriate separation of both the compounds with toluene/ethyl acetate/formaldehyde (6:3:1; v/v) (Srivastava et al. [2008]), no identification of vanillin peak using hexane/ethyl acetate (5:2; v/v) (Hennig et al. [2011]) and 1-butanol/acetic acid/water (66:17:17; v/v) (Males et al. [2001]). Chloroform/methanol/formic acid (85:15:1; v/v) with spraying reagent 2, 4-dinitrophenylhydrazine was used for vanillin detection but resulted in poor peak visualization and separation of vanillin (Mabinya et al. [2006]). Sharp peaks were obtained with a presaturation of mobile phase for 20 min. FA and vanillin was quantified using UV detector and scanning at 320 and 312 nm. The peaks corresponding to FA and vanillin in samples had the same retention time when compared to their respective standards.

Linearity

Linearity was achieved with concentrations ranging from 50 to 250 ng/band for FA and 50 to 200 ng/band for vanillin (Figures 1 and 2). The regression equation and correlation coefficient of FA was found to be y = 37.94 × X + 621.04, R2 = 0.9975 with SD = 1.95% and for vanillin y = 36.016 × X + 1128.3, R2 = 0.9991 with SD = 1.60%.

Figure 1
figure 1

Standard graph of FA by HPTLC method.

Figure 2
figure 2

Standard graph of vanillin by HPTLC method.

Precision

To get accurate chromatographic results, the precision of the chromatographic method must be analyzed and confirmed whether it is fit for purpose which is adequate to the analytical requirements and it is evaluated in terms of intra- and inter-day precision. The standard deviation and percent relative standard deviation will evaluate the variation limit of the analysis. The value lower than 2% indicate the method is more precise to variation which assumes that the chromatograph does not malfunction after the system precision testing has been performed (Indrayanto [2011]). Peak areas measurement of 100, 150, and 200 ng concentration of FA and vanillin showed %RSD less than 2 as shown in Table 1. Both intra- and inter-day results suggested an excellent method of precision which ensures the objective of the method development phase to be reproducible.

Table 1 Intra- and inter-day precision ( n= 5)

Recovery

Recovery is an important parameter as it offers information about the recovery of the analyte from the sample preparation and the effect of matrix. If the recovery is close to 100% then it implies that the proposed analytical method is free from constant and proportional systematic error (Srivastava [2011]). Recovery of FA at three different levels obtained was 98.47%, 99.1%, and 98.75% with 98.77% average recovery and of vanillin was 96.51%, 99.19%, and 99.65% with 98.45% average recovery. The results are shown in Table 2. The percentage recovery was close to 100% which indicated no interference of any other compound and representing the accuracy of the method. This indicates the suitability of the method for the routine analysis.

Table 2 Recovery of FA and vanillin

Specificity

Specificity is a method which provides a response for only a single analyte. This study is performed to check how accurately and specifically the analyte of interest is estimated in the presence of other components with system interference during detection and quantification of analyte (Indrayanto and Yuwono [2010]; Kakde et al. [2008]). Retention factor values of standard FA and vanillin were compared with sample extract. There was no interference of other peaks as shown in Figures 3, 4, and 5. The standard and sample was compared with respect to peak start, peak apex, and peak end of the bands which showed the specificity between extracts and standard compound and purity of the peaks (Table 3). Overlay of the standard and sample compound using spectral scan was compared to confirm the specificity (Figures 6 and 7). Spectral scan of both the compounds showed maximum absorption at 320 nm for ferulic acid and 312 nm for vanillin. Thus, results obtained from comparison of peak and spectral scan showed the method is specific for detection of FA and vanillin.

Figure 3
figure 3

HPTLC chromatogram of standard FA.

Figure 4
figure 4

HPTLC chromatogram of standard vanillin.

Figure 5
figure 5

HPTLC chromatogram of FA and vanillin present in crude extract.

Table 3 Specificity of FA and vanillin
Figure 6
figure 6

Overlay of UV absorption spectra of standard FA with crude extract.

Figure 7
figure 7

Overlay of UV absorption spectra of standard vanillin with crude extract.

Robustness

Robustness can be described as the ability to reproduce the analytical method under different circumstances and provides an indication of its reliability during normal phase. It was introduced to avoid problems in inter-laboratory studies and to discover the potentially repairable factors (Van der Hyden et al. [2001]). According to the ICH guidelines, the evaluation of robustness should be considered during the development phase and depends on the type of procedure under study. For robustness evaluation by chromatographic method, the acceptance criteria are that, the Rf values of all standards should lie within the acceptance criteria of the precision method by performing variations in considered parameters. Calculations of standard or relative error are common ways to look at the data and departures from deviations in the data will directly affect robustness. Variations made in mobile phase composition, mobile phase volume, saturation time, and temperature showed less %RSD and SD for FA and vanillin by spotting 150 ng/spot (n = 3) (Table 4). No significant change in Rf or response to FA and vanillin was observed which indicated robustness of the method. Robustness study has provided valuable information about the quality and reliability of the method and no further development or optimization is necessary.

Table 4 Robustness of FA & Vanillin

Limit of detection and limit of quantification

LOD is based on the analyte response sensitivity (response per amount or concentration per time, using either peak height or area), and LOQ is the lowest amount of analyte that can be quantitatively determined in sample with defined precision and accuracy under standard conditions. LOQ is usually a multiple of LOD (Kakde et al. [2008]). Standard deviation and slope used in the equations are used to determine LOD and LOQ which is equivalent to instrument sensitivity for the specific analyte, reinforcing that the LOD/LOQ details, are expressed in units of analyte concentration (Apostol et al. [2012]). It is important to determine sensitivity using analyte amounts near to their detection limits. LOD and LOQ with signal-to-noise ratio of FA were found to be 13.63 and 45.42 ng/band, respectively. LOD and LOQ of vanillin were found to be 10.59 and 35.30 ng/band, respectively. Sensitivity of the method is evaluated with regard to LOD and LOQ (Kakde et al. [2008]). These results indicate the sensitivity of the method which can be used for quantification of the compound.

System suitability

System suitability test (SST) is an integral part of many analytical procedures. The test is based on the conception that, the equipment, electronics, analytical operations, and samples to be analyzed represent an integral system that can be evaluated. System suitability test parameters to be established for a particular procedure depend upon the type of procedure being validated. It is the ability of the analytical method to detect analyte quantitatively in the presence of other components which are expected to be present in the sample and they should be chromatographed along with the analyte to check the system suitability and retention factor of the required analyte (Dolan [2004]). On the basis of repeatability relative standard deviations of peak response, SST was analyzed. Retention factor of FA and vanillin were 0.48 and 0.56, respectively. Standard deviation of FA was 2.95 ± 0.13 with 0.039% RSD and standard deviation of vanillin was 3.62 ± 0.12 with 0.056% RSD. The low %RSD indicates the reproducibility and the system suitability of the method.

Conclusion

The simultaneous detection of FA and vanillin by the HPTLC method can be performed as there is noteworthy difference in their retention factor values. The proposed method was developed and validated by ICH guidelines which is simple, rapid, accurate, precise, sensitive, and eco-friendly. The mobile phase selected was toluene/1, 4-dioxan/acetic acid 9:2.5:0.4 (v/v) which resulted in proper peak separation as compared to different mobile phase which were not able to separate the compounds properly. Other chromatographic methods like HPLC, GC, and spectrophotometric mentioned may not be applied for routine check because of the various shortcomings as compared to HPTLC. The total optimized method is therefore useful in both qualitative and quantitative analysis for routine assays in pharmaceutical and food industry within acceptable limits.

References

  1. Apostol I, Krull I, Kelner D: Analytical Method Validation for Biopharmaceuticals. Analytical Chemistry 2012. doi:[10.5772/52561]

    Google Scholar 

  2. Bakshi M, Singh SJ: Development of validated stability-indicating assay methods-critical review. J Pharm Biomed Anal 2002, 28: 1011–1040. 10.1016/S0731-7085(02)00047-X

    Article  CAS  Google Scholar 

  3. Butehorn U, Pyell U: Micellar electrokinetic chromatography as a screening method for the analysis of vanilla flavorings and vanilla extracts. J Chromatogr A 1996, 736: 321–332. 10.1016/0021-9673(95)01335-0

    Article  Google Scholar 

  4. Chiou A, Karathanos VT, Mylona A, Salta FN, Preventi F, Andrikopoulos NK: Currants ( Vitisvinifera L .) content of simple phenolic and antioxidant activity. Food Chem 2007, 102: 516–522. 10.1016/j.foodchem.2006.06.009

    Article  CAS  Google Scholar 

  5. Constantine DS: Extraction, separation, and detection methods for phenolic acids and flavonoids. J Sep Sci 2007, 30: 3268–3295. 10.1002/jssc.200700261

    Article  Google Scholar 

  6. Dolan JW: System suitability. LC troubleshooting, BASi Northwest Laboratory, McMinnville, Oregon, USA 2004, 17(6):328–332.

    Google Scholar 

  7. 6th edn. Strasbourg, Council of Europe; 2008.

  8. Graf E: Antioxidant potential of ferulic acid. Free Radical Biol Med 1992, 13: 435–448. 10.1016/0891-5849(92)90184-I

    Article  CAS  Google Scholar 

  9. Gross B, Asther M, Corrieu G, Brunerie P: Production of vanillin by bioconversion of benzenoid precursors by Pycnoporus. U.S. Patent 5262315 B. 1993.

    Google Scholar 

  10. Hennig L, Garcia GM, Giannis A, Bussmann RW: New constituents of Baccharis genistelloides (Lam.). Pers. Arkivoc 2011, 6: 74–81.

    Google Scholar 

  11. Ho K, Yazana LS, Ismail N, Ismail M: Apoptosis and cell cycle arrest of human colorectal cancer cell line HT-29 induced by vanillin. Cancer Epidem 2009, 33: 155–160. 10.1016/j.canep.2009.06.003

    Article  CAS  Google Scholar 

  12. Huck CW, Stecher G, Scherz H, Bonn G: Analysis of drugs, natural and bioactive compounds containing phenolic groups by capillary electrophoresis coupled to mass spectrometry. Electrophoresis 2005, 26: 1319–1333. 10.1002/elps.200410315

    Article  CAS  Google Scholar 

  13. Indrayanto G: Analytical Aspects of High Performance Thin Layer Chromatography. MM. Srivastava edn, High-Performance Thin-Layer Chromatography (HPTLC), Springer-Verlag Berlin Heidelberg; 2011.

    Book  Google Scholar 

  14. Indrayanto G, Yuwono MS: TLC: Validation of Analyses. In Cazes J Encyclopedia of Chromatography, 3rd edn. Taylor & Francis Group, London; 2010.

    Google Scholar 

  15. Kakde RB, Kotak VH, Kale DL: Estimation of bisoprolol fumarate in pharmaceutical preparations by HPTLC. Asian J Res Chem 2008, 1(2):70–73.

    CAS  Google Scholar 

  16. Krishna Veni N, Meyyanathan SN, Reddy AA, Sompura SA, Elango K: Analysis of Vanillin In Food Products By High Performance Thin Layer Chromatography. J Adv Sci Res 2013, 4(1):48–51.

    CAS  Google Scholar 

  17. Lomascolo A, Stentelaire C, Asther M, Lesage-Meessen L: Basidiomycetes as new biotechnological tools to generate natural aromatic flavours for the food industry. Trends Biotechnol 1999, 17: 282–289. 10.1016/S0167-7799(99)01313-X

    Article  CAS  Google Scholar 

  18. Mabinya LV, Mafunga T, Brand JM: Determination of ferulic acid and related compounds by thin layer chromatography. Afr J Biotechnol 2006, 5: 1271–1273.

    CAS  Google Scholar 

  19. Mabry TJ, Markham KR, Thomas MB: The Systematic Identification of Flavonoids. Springer, New York, NY, USA; 1970.

    Book  Google Scholar 

  20. Macheix JJ, Fleuriet A, Billot J: Fruit Phenolics. CRC Press, Bota Raton, FL; 1990.

    Google Scholar 

  21. Males Z, Medic-Saric M: Optimization of TLC analysis of flavonoids and Phenolic acids of Helleborus atrorubens Waldst. et Kit. J Pharm Biomed Anal 2001, 24(3):353–9. 10.1016/S0731-7085(00)00455-6

    Article  CAS  Google Scholar 

  22. Olsson L, Samuelson O: Chromatography of aromatic acids and aldehydes and phenols on cross-linked polyvinylpyrrolidone. J Chromatogr 1974, 93: 189–199. 10.1016/S0021-9673(00)83030-7

    Article  CAS  Google Scholar 

  23. Priefert H, Rabenhorst J, Steinbuchel A: Biotechnological production of vanillin. Appl Microbiol Biotechnol 2001, 56: 296–314. 10.1007/s002530100687

    Article  CAS  Google Scholar 

  24. Prinjaporn TN, Namthip N, Poomrat R, Weena S, Orawon C: Simple and Rapid Determination of Ferulic Acid Levels in Food and Cosmetic Samples Using Paper-Based Platforms. Sensors 2013, 13: 13039–13053. 10.3390/s131013039

    Article  Google Scholar 

  25. Raisi A, Aroujaliana A, Kaghazchia T: Multicomponent pervaporation process for volatile aroma compounds recovery from pomegranate juice. J Membr Sci 2008, 322: 339–348. 10.1016/j.memsci.2008.06.001

    Article  CAS  Google Scholar 

  26. Rakchoy S, Suppakul P, Jinkarn T: Antimicrobial effects of vanillin coated solution for coatingpaperboard intended for packaging bakery products. As J Food Ag-Ind 2009, 2: 138–147.

    Google Scholar 

  27. Rao SR, Ravishankar GA: Biotransformation of isoeugenol to vanilla flavour metabolites and capsaicin in suspended and immobilized cell cultures of Capsicum frutescens: study of the influence of β-cyclodextrin and fungal elicitor. Process Biochem 1999, 35: 341–348. 10.1016/S0032-9592(99)00077-1

    Article  CAS  Google Scholar 

  28. Renger B: Quantitative planar chromatography as a tool in pharmaceutical analysis. J AOAC Int 1993, 76: 7–13.

    CAS  Google Scholar 

  29. Renger B: Contemporary thin layer chromatography in pharmaceutical analysis. J AOAC Int 1998, 81: 333–339.

    CAS  Google Scholar 

  30. Sharma UK, Sharma N, Gupta AP, Kumar V, Sinha AK: RP-HPTLC densitometric determination and validation of vanillin and related Phenolic compounds in accelerated solvent extract of Vanilla planifolia . J Sep Sci 2007, 30: 3174–3180. 10.1002/jssc.200700229

    Article  CAS  Google Scholar 

  31. Sostaric T, Boyce MC, Spickett EE: Analysis of volatile components in vanilla extract and flavorings by solid-phase microextraction and gas chromatography. J Agric Food Chem 2000, 48: 5802–5807. 10.1021/jf000515+

    Article  CAS  Google Scholar 

  32. Srivastava MM: An Overview of HPTLC. In A Modern Analytical Technique with Excellent Potential for Automation, Optimization, Hyphenation, and Multidimensional Applications. High-Performance Thin-Layer Chromatograpy (HPTLC), Springer-Verlag Berlin Heidelberg; 2011.

    Chapter  Google Scholar 

  33. Srivastava SK, Singh AP, Rawat AKS: Pharmacognostical and phytochemical evaluation of Lycopodium clavatum stem. J Sci Ind Res 2008, 67: 228–232.

    CAS  Google Scholar 

  34. Takahashi M, Sakamaki S, Fujita A: Simultaneous analysis of Guaiacol and vanillin in a vanilla extract by using High Performance Liquid Chromatography with Electrochemical detection. Biosci. Biotechnol. Biochem 2013, 77(3):595–600. 10.1271/bbb.120835

    Article  CAS  Google Scholar 

  35. Tilay A, Bule M, Kishenkumar J, Annapure U: Preparation of ferulic acid from agricultural wastes: its improved extraction and purification. J Agric Food Chem 2008, 56: 7644–8. 10.1021/jf801536t

    Article  CAS  Google Scholar 

  36. Tilay A, Bule M, Annapure U: Production of Biovanillin by One-Step Biotransformation Using Fungus Pycnoporous cinnabarinus . J Agric Food Chem 2010, 58: 4401–4405. 10.1021/jf904141u

    Article  CAS  Google Scholar 

  37. Van der Hyden Y, Nijhuis A, Smayers-Verbeke J, Vandeginste BMG, Massart DL: Guidance for robustness/ruggedness test in method validation. J Pharm Biomed Anal 2001, 24: 723–753. 10.1016/S0731-7085(00)00529-X

    Article  Google Scholar 

  38. Wu SL, Chen JC, Li CC, Lo HY, Ho TY, Hsiang CY: Vanillin Improves and Prevents Trinitrobenzene Sulfonic Acid-Induced Colitis in Mice. J Pharmacol Exp Ther 2009, 330: 370–376. 10.1124/jpet.109.152835

    Article  CAS  Google Scholar 

  39. Xu F, Sun RC, Sun JX, Liu CF, Heb BH, Fan JS: Determination of cell wall ferulic and p-coumaric acids in sugarcane bagasse. Anal Chim Acta 2005, 552: 207–217. 10.1016/j.aca.2005.07.037

    Article  CAS  Google Scholar 

  40. Zheng L, Zheng P, Sun Z, Bai Y, Wang J, Guo X: Production of vanillin from waste residue of rice bran oil by Aspergillus niger and Pycnoporus cinnabarinus . Bioresour Technol 2007, 98: 1115–1119. 10.1016/j.biortech.2006.03.028

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Uday S Annapure.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

SH and SD has performed all the experimental and analytical work and drafted the manuscript. The guidelines for all the mentioned part was provided by UA. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Cite this article

Hingse, S.S., Digole, S.B. & Annapure, U.S. Method development for simultaneous detection of ferulic acid and vanillin using high-performance thin layer chromatography. J Anal Sci Technol 5, 21 (2014). https://doi.org/10.1186/s40543-014-0021-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40543-014-0021-6

Keywords